YOKOGAWA SG750 Stack Gas Analyzer

YOKOGAWA SG750 Stack Gas Analyzer

The SG750 Stack Gas Analyzer consists of an infrared gas analyzer, a zirconia oxygen analyzer, and a sampling unit.
It can simultaneously measure up to 5 components: NOx, SO2, CO2, CO, and O2.

The use of interference compensation in the analyzing section virtually eliminates the effects of interference from other components. The SG750 features many functions including autocalibration, conversion to O2 values, averaging, and high/low alarms. It is used for measuring and monitoring emissions from various types of boilers and waste incinerators.

 Bulletin for SG750  

 General Specification for SG750  

Features

*

Simultaneously and continuously measures concentrations of up to 5 gas components NOX, SO2, CO, CO2 and O2.

*

The use of interference compensation in the analyzing section virtually eliminates the effect of concentration of any gas component on the concentration readings of others.

*

Adopts a double beam infrared analyzer, a Zirconia oxygen analyzer meter or paramagnetic oxygen meter with high sensitivity and excellent long-term stability.

*

The maintenance type on the front saves space.

*

High sensitivity and wide dynamic range, allowing for switched ranging of up to 1:25 and range can be changed.

Specification

Measurement

Up to 4 components (NOx, SO2, CO2, CO) and O2 in flue gases.

Measurering method

NOx, SO2, CO2, CO: Non-dispersive Infrared method

O2: Zirconia or paramagnetic method

Measuring ranges

NOx:

:0-50 ppm to 0-5000 ppm

SO2:

0-100 ppm to 0-1000 ppm

CO2:

0-1 vol% to 0-20 vol%

CO:

0-50 ppm to 0-5000 ppm

O2:

0-10/0-25 vol%

Display

LCD with backlight

Output signal

5 points, 4-20mA DC or 0 - 1 V DC

Contact output

Each component range identification, analyzer failure, calibration failure, calibration status, maintenance status, CO peak count alarm, each instantaneous value alarm, and analyzer power shutdown

Contact input

Voltage-free contact, auto calibration start, average reset, Range switching output hold, remote pump off

Automatic calibration function

Zero, span are auto calibrated (calibration cycle settable)

 • 

Interval range;1 to 99hours (1hour increments) or 1 to 40days (1day increments)

 • 

Time of calibration gas flow: 60 to 599seconds (1second increments)

Construction

Outdoor/indoor stand-alone system, non-explosionproof, rainproof, single swing front door, standard plate thickness of 2.3 mm (both housing and door)

Ambient temperature

-5 to 40

Ambient humidity

90%RH or less

Power supply

100/110/115/200/230 V AC ±15%, 50/60 Hz ±0.5 Hz

Power consumption

Max. 600 VA

Weight

Approx. 300 kg (without standard gases)

Sample gas conditions

Temperature: 1400 or less

Dust: 500 mg/Nm3 or less

Pressure: -1 to 5kPa, -3 to 3kPa, -5 to 1kPa

Flow rate: Approx. 2 l/min

Performance

Linearity

±1.0% F.S.

Repeatability

±0.5% F.S.

Drift

Zero:

±1% F.S./week, ±2% F.S./ week for the range of 200 ppm or less

±2% F.S./month for zirconia oxygen analyzers

Span:

±2% of F.S./week

±2% F.S./month for zirconia oxygen analyzer

Response time
(From the inlet of
the system)

Approx. 4 minutes for SO2

Approx. 2 minutes for others

#