|
Town Gas Application |
|
|
Purpose |
Measurement and control of the calorific value of town gas. |
|
Measurement |
WI or calorific value of fuel gas. |
|
Measuring Range |
3 to 62 MJ/Nm3 |
|
Conditions at the Sampling Point |
|
Dust: 5 mg/Nm3 or less |
|
Temperature: 50℃ or less |
|
Humidity: dew point of 0℃ or less |
|
Pressure: (1) 10 to 20 kPa: standard, (2) 10 kPa or under: with pump, (3) 100 to 600 kPa: with pressure reducing valve | |
|
Range |
Select scale range (Span): General Gas: 30 to 50% of maximum value of the span. Butane or Butene + Air:20 to 30% of maximum value of the span. Propane or Propylene + Air: 25 to 40% of maximum value of the span. |
|
Output |
1 to 5 V DC, 4 to 20 mA DC (simultaneously), non-isolated, load resistance 750Ω or less |
|
Alarm Contact Output |
Flame off alarm; 100 V AC, 5 A, closed when alarm occurs (resistance load) Temperature alarm; 100 V AC, 3 A, closed when alarm occurs (resistance load) |
|
Contact Input |
Remote ignition (Tokuchu); 24 V DC, 0.1A or more |
|
Sample Gas Flow Rate |
Approx. 10 l/min |
|
Utility |
Instrument Air: Approx. 50 Nl/min, pressure 300 to 700 kPa, dew point of 0℃ or less Power Supply: 100 V AC ±10%, single phase, 50/60 Hz, 860 VA max. |
|
Panel |
Construction: For indoor installation, rack panel. |
|
Ambient Temperature |
0 to 40℃ (Little temperature variation, particularly no rapid change in temperature, allowed) |
|
Repeatability |
When the calorific value means 6.3 MJ/Nm3 or higher, Calorific : ±1.0% of measured value WI : ±0.5% of measured value |
|
Steel Mill Application |
|
|
|
|
Purpose |
Measurement and control of the calorific value of fuel gas for a steel mill. |
|
Measurement |
WI or calorific value of fuel gas. |
|
Measuring Range |
3 to 62 MJ/Nm3 |
|
Conditions at the Sampling Point |
|
Dust: 100 mg/Nm3 or less |
|
Temperature: 50℃ or less |
|
Humidity: dew point of 0℃ or less |
|
Pressure: (1) 8 to 15 kPa: standard, (2) 8 kPa or under: with pump | |
|
Range |
Select scale range (Span): General Gas: 30 to 50% of maximum value of the span. Butane or Butene + Air:20 to 30% of maximum value of the span. Propane or Propylene + Air: 25 to 40% of maximum value of the span. |
|
Output |
1 to 5 V DC, 4 to 20 mA DC (simultaneously), non-isolated, load resistance 750Ω or less |
|
Alarm Contact Output |
Flame off alarm; 100 V AC, 5 A, closed when alarm occurs (resistance load) Temperature alarm; 100 V AC, 3 A, closed when alarm occurs (resistance load) |
|
Contact Input |
Remote ignition (Tokuchu); 24 V DC, 0.1A or more |
|
Sample Gas Flow Rate |
Approx. 10 l/min |
|
Utility |
Water: Approx. 0.2 l/min, pressure 200 to 600 kPa Instrument Air: Approx. 50 Nl/min, pressure 300 to 700 kPa, dew point of 0℃ or less Power Supply: 100 V AC ±10%, single phase, 50/60 Hz, 1100 VA max. |
|
Panel |
Construction: For indoor installation, rack panel. |
|
Ambient Temperature |
0 to 40℃ (Little temperature variation, particularly no rapid change in temperature, allowed) |
|
Repeatability |
When the calorific value means 6.3 MJ/Nm3 or higher, Calorific : ±1.0% of measured value WI : ±0.5% of measured value |