YOKOGAWA GD402, GD40 Gas Density Analyzers |
YOKOGAWA GD402, GD40 Gas Density Analyzers
There are many applications where the accurate, repeatable measurement of gas density is essential. From blending of fuel, to product quality, to Hydrogen purity and replacement in utility turbines, Yokogawa’s family of Gas Density products allow you to make accurate measurements with superior reliability and ease of maintenance and operability.
The GD402 gas density analyzer and GD40 detector not only provide continuous measurement of gas density, but also several other valuable parameters, including specific gravity and molecular weight.
The GD402 is a rugged microprocessor-based converter designed in two versions to meet both general area and explosion-proof application requirements. In addition to the display of several key data items, the converter also provides the choice of three different means for calibration: automatic,semi-automatic and one-touch manual operation.
The GD40 detector is designed for intrinsically safe and flameproof, explosion protected applications. It is designed to be virtually maintenance free for all accepted applications. |
Features
* |
Proven design Highly responsive and sensitive measurement of density. Specific gravity, molecular weight and gas concentration can also be displayed using Yokogawa's gas density analyzing techniques. |
* |
Low installation cost Both explosionproof and general purpose converters are designed for easy mounting on a pipe. Wiring between the detector and converter is based on a two-wire system,keeping installation cost to a minimum |
* |
Detector features
・ |
Resistant to external vibrations |
・ |
Outstanding stability against sudden changes in gas temperature (within 1 g/m3 for sudden changes in gas temperature of 10℃) |
・ |
The multi-mode self-oscillation circuit minimizes drift caused by the sensor itself or by oil mist, dust, moisture,etc. sticking to the sensor |
・ |
Easy cleaning and regeneration of sensor. Should the sensor be contaminated with dust and/or mist, then it can be easily cleaned and returned to its original condition. |
・ |
Only routine maintenance is required (for example, once every 3 months). | |
Specification
Measuring object |
Gas density of all gases except for corrosive and acetylene gas |
Measuring principle |
Vibration type (resonant frequency) |
Sample gas conditions |
Temperature: -10 to 50℃ (non-condensing) |
Pressure: Max. 588.4 kPa (abs) |
Gas flow:0.1 to 1 l/min | |
Measuring range |
Gas density: 0 to 6 kg/m3 |
Output signal |
Output 1: 4-20 mA DC, Isolated, load resistance: 600Ω max. |
Output 2: 4-20 mA DC, Isolated, load resistance: 600Ω max. | |
Explosionproof approval Converter |
Rainproof for outdoor use, equivalent to IP65 / NEMA 4X Ambient temperature :-10 to 55℃ Ambient humidity : 5 to 95%RH
GD402T: |
FM Explosionproof Approval. Explosionproof for Class I, Division 1, Groups B, C and D; Dust Ignitionproof for Class II, III, Division 1, Groups E, F and G. Enclosure : NEMA Type 4X Temperature code : T6 Electrical connection: 1/2NPT female |
GD402V: |
CSA Explosionproof Approval. Explosionproof for Class I, Division 1, Groups B, C and D; Dust Ignitionproof for Class II, III, Division 1, Groups E, F and G. Enclosure : Type 4X Temperature code : T6 Electrical connection: 1/2NPT female |
GD402R:: |
TIIS Explosioproof approval. Explosionproof code : Exd IIB+H2T6 Temperature code : T6 Electrical connection: G3/4 female | |
Explosionproof approval detector |
Rainproof for outdoor use, equivalent to IP65 / NEMA 4X Ambient temperature :-10 to 60℃ Ambient humidity : 5 to 95%RH
GD40T: |
FM Explosionproof and Intrinsically safe Approval. Explosionproof for Class I, Division 1, Groups B, C and D; Dust Ignitionproof for Class II, III, Division 1, Groups E, F and G with Intrinsically Safe sensor for Class I, II, III, Division 1, Groups B, C, D, E, F and G. Enclosure : NEMA Type 4X Temperature code: T5 Electrical connection : 1/2NPT female Process connection : 1/4NPT female |
GD40V: |
CSA Explosionproof and Intrinsically safe Approval. Explosionproof for Class I, Division 1, Groups B, C and D; Dust Ignitionproof for Class II, III, Division 1, Groups E, F and G with Intrinsically Safe sensor for Class I, II, III, Division 1, Groups B, C, D, E, F and G. Enclosure : Type 4X Temperature code : T5 Electrical connection : 1/2NPT female Process connection : 1/4NPT female |
GD40R: |
TIIS Explosionproof and Intrinsically safe Approval. Explosionproof code : Exd [ia] IIB+H2T5 Temperature code : T5 Electrical connection : G1/2 female Process connection : Rc1/4 female | |
Power supply |
100 to 240 V AC, 24V DC, 50/60 Hz |
Performance
Linearity |
±1% F.S. |
Repeatability |
±0.001 kg/m3 or ±0.5% F.S. whichever is greater |
Long term stability |
±0.003 kg/m3 per month |
Response time 90% |
Approx. 5 sec | |
|