YOKOGAWA IR400 Infrared Gas Analyzer

YOKOGAWA IR400 Infrared Gas Analyzer

The IR400 infrared gas analyzer is capable of measuring the concentrations of NO, SO2, CO2, CO, CH4 and O2 components in sample gas. NO, SO2, CO, CO and CH4 are measured by the nondispersion infrared method (NDIR), while O2 is measured by the built-in type paramagnetic method sensor or an external mount-type zirconia method sensor. A maximum of 5 components including O2 (max. 4 components except for O2 measurement) are simultaneously measurable.

The mass flow type twin detector of high sensitivity and reliability adopted in the infrared ray method detection unit makes the measurement hardly affected by interfering components. In addition, the IR400 includes a microprocessor and has a large-size liquid crystal display, iproviding easy operation, high accuracy and multiple functions.

 General Specification for IR400

 User's Manual for IR400 

 Click here to request a technical document for Infrared Gas Analyzers

Features

*

Simultaneous and continuous measurement of up to 5 components including O2 and 4 components selected from among NO, SO2, CO, CO2, and CH4.

*

The mass flow type twin detector of high sensitivity and reliability minimizes interference from other gas components, ensuring excellent stability.

*

Incorporating O2 correction, average value computing, automatic calibration, one touch calibration, upper/lower limit alarm, remote measurement range changeover, range identification signal output, etc., the analyzer accommodates different application requirements.

*

Large LCD provides clear indications of all measured components and computed values and easy interactive operation.

*

Unitized construction of the main body on the 19-inch rack and of the signal input/output terminal module allows easy configuration of a gas analyzer system.

*

A maximum range ratio of 1:25 is achieved.

Specifications

Measurement principle

NO, SO2, CO2, CO, CH4: Non-dispersive infrared method, single light source-double beams
O2: Paramagnetic type (built-in), or zirconia type (external)

Sample gas conditions

Flow rate: 0.5 ±0.2 L/min

Temperature: 0 to 50

Pressure: 10 kPa or less (Gas outlet side should be open to the atmospheric.)

Dust: 100 μg/Nm3 or less in particle size of 1 mm or less

Mist: Unallowable

Moisture: Below a level where saturation occurs at 2 (condensation unallowable).

Corrosive component: HCl 1 ppm or less

Measurable gas components and measuring range

Minimum range

Maximum range

NO

0-50ppm

0-5000ppm

SO2

0-50ppm

0-10vol%

CO2

0-20ppm

0-100vol%

CO

0-50ppm

0-100vol%

CH4

0-500ppm

0-100vol%

O2(built in)

0-5vol%

0-25vol%

O2 (External Zirconia)

0-5vol%

0-25vol%

Analog output signals

4 to 20mA DC or 0 to 1V DC, nonisolated output

Analog input signal

For signal input from externally installed O2 sensor. Nonisolated input.

Contact input

Non-voltage contact (ON/0V, OFF/5VDC, 5mA flowing at ON) Isolated from the internal circuit with photocoupler. Contact inputs are not isolated from one another.

Transmission output

Solenoid valve drive signal for automatic calibration. Transistor output (100mA or less)

Ambient temperature

-5 to 45

Ambient humidity

90% RH max., noncondensing

Dimensions
(H x W x D)

Analyzer main unit; 177 x 483 x 600 mm
Input/output terminal module; 164 x 318 x 55 mm

Weight

Approx. 22kg (Only Analyzer)

Power supply

85 to 264 V AV, 50/60 Hz (3-pin inlet terminal used)

Performance

Linearity

±1% F.S

Repeatability

±0.5% F.S.

Zero drift

±1% F.S. / Week

Span drift

±2% F.S. / Week

Response time
(for 90% FS response)

Within 60seconds including replacement time of sample gas
(when gas flow rate is 0.5 l/min)

Temperature drift

±1.5% F.S. / 10

Effects of measured gas flow rate

±1% F.S. for the rated flowrate ±10%

Applications

1. Measurement system for combustion exhaust gas from refuse incinerator and boiler
2. Gas from various types of industrial furnaces

#