YOKOGAWA IR100 Universal Infrared Gas Analyzers

YOKOGAWA IR100 Universal Infrared Gas Analyzers

The IR100TA and IR100A, dedicated for analysis of either CO2, CO, or CH4, IR100TB and the IR100B, dedicated for simultaneous analysis of CO2 and CO, are available. The analyzers are easy to maintain and works steadily over long periods.

The ideal application of these analyzers are continuous measurement for the control and monitoring of combustion in various industrial furnaces as well as continuous measurement for the study of plants.

 Bulletin for IR100

 General Specification for IR100

 User's Manual for IR100

 Click here to request a technical document for Infrared Gas Analyzers

Features

*

Two available models: simultaneous dual-gas (CO and CO2) measurement model and single-gas dual-range measurement model

*

Stable operation over long periods

*

Influenced less by presence of another interfering gas

*

Easy maintenance

*

Wide variety of additional functions: self-diagnosis, simple calibration, automatic calibration, remote switching measuring range, and range identification

Specifications

Models

Horizontal model :

IR100A (single-gas, dual-range analyzer),
IR100B (dual-gas, dual-range analyzer)

Vertical model :

IR100TA (single-gas, dual-range analyzer),
IR100TB (dual-gas, dual-range analyzer)

Measuring principle

Non-dispersive infrared absorption, deflection method, single infrared-source, single-flux (single-beam)

Sample gas conditions

Flowrate: 0.5 to 1 l/min
Pressure: 500 Pa or more
Temperature: 0 to +50

Dust: 100μg/Nm3 with the particle size of 1μm or less
Mist: none
Humidity: Avoid concentration.
Corrosive gases sampled:
 NOx; 1000 ppm or less
 SO2; 1000 ppm or less
 HCl; 1 ppm or less
 Others; None

Measuring gas component and measuring range

IR100A, IR100TA; single-gas dual-range analyzers
CO2: 0-500ppm to 0-100%
CO: 0-500ppm to 0-100%
CH4: 0-1000ppm to 0-100%
IR100B, IR100TB; dual-gas dual-range analyzers
CO2: 0-500ppm to 0-100%
CO: 0-500ppm to 0-100%

Output signals

Output 1: 0 to 1V DC, non-isolated, linearized output signals
Output 2: 4 to 20mA (maximum allowable load resistance: 550 Ω), simultaneous, non-isolated, linearized signals

Ambient temperature

-5 to 45

Ambient humidity

90%RH or less

Installation

Indoor

Weights

Approx. 12kg

Power supply

100V AC, 115 V AC, 220 V AC(±10)50/60Hz

Performance

Linearity

±-2% FS

Repeatability

±0.5% FS for the primary range (lower range)
±1% FS for the secondary range (higher range)

Zero drift

±2% FS/week

Span drift

±2% FS/week

Response time
(90% response)

15 seconds or less including the time required for substitution of the sample cell

Applications

Blast furnace

CO; 0 to 40%   CO2; 0 to 30%, 0 to 40%, 0 to 50%

Converter

CO; 0 to 100%   CO2; 0 to 100%

Electric furnace

CO; 0 to 100%   CO2; 0 to 100%

Coke oven

CO; 0 to 100%   CO2; 0 to 100%

Cement kiln

CO; 0 to 1%, 0 to 5%

Coal kiln

CO; 0 to 1%, 0 to 5%

Carbonizing furnace

CO2; 0 to 1%, 0 to 2%

Transforming furnace

CO2; 0 to 0.5%, 0 to 1%

#