HT INDUSTRIAL AUTOMATION LIMITED

 
YOKOGAWA GC8000 Process Gas Chromatograph

YOKOGAWA GC8000 Process Gas Chromatograph

The GC8000 has a built-in 12-inch color touchscreen display that dramatically simplifies maintenance. At the touch of the screen, the technician can access all of the analytical parameters and measurement results; displayed in easy-to-understand graphical color screens.

The GC8000 also brings a number of important innovations in its gas chromatograph analytical capabilities. For the first time, parallel chromatography is made practical with the introduction of the GC Module (GCM) concept. By setting up virtual GCs within a single analyzer, all chromatograph settings, displays, and data are truly segregated for easy understanding and maintenance. There are even built-in graphical overview screens showing each of the individual GC Modules.

The GC8000 uses the same proven analytical hardware found in our previous model of GC; recognized for its reliable and precise performance. But with the GC8000, the analytical possibilities are greatly expanded through the multiple oven capability of the analyzer.

The communications network of the GC8000 is based on the industry standard Ethernet structure to provide flexible yet secure transmission of data to GC maintenance workstations and the plant DCS system. The GC8000 can be set up for either a single Ethernet network or a redundant network with two completely isolated Ethernet networks if desired.

 Bulletin for GC8000 

 General Specification for GC8000

Features

* High-speed analysis with multi-oven structure
* Improved operation with a large touch panel
* High reliability
* Improved functionality

Specifications

Measurable object

Gas or volatile liquid (400 or lower boiling point)

Analysis method

Gas chromatography

Measurable range

Depends on analysis conditions
TCD      : 1ppm to 100%
FID       : 1ppm to 100%
FID (with methanizer): 1 ppm to 0.1%
FPD      : 1ppm to 0.1%

Number of components to be measured

Maximum of 999 (total number of components in all streams including calibration standard sample streams)

Analysis period

Maximum of 21600.0 seconds (six hours)

Number of streams to be measured

Maximum of 31 (including standard sample streams)

Material of sample-contact parts

RV: 316SS, Hastelloy-C, Rulon, PTFE (Teflon, Bearee)
LSV: 316SS, Hastelloy-C, Rulon, Glass, PTFE (Teflon, Bearee), Fluororubber (Viton), perfloroelastomer (Kalrez)

Repeatability

Depends on analysis conditions
Gas sample: ±1% of full scale for measuring ranges (2σ)
Liquid sample: ±2% of full scale for measuring ranges (2σ)

*

The value may vary depending on the specifications and conditions. For details, contact Yokogawa.

Analyzer specifications

Type of protection

Pressurized enclosure and flameproof enclosure

Area classification

FM:
Type X Pressurization and Explosionproof for Class I, Division 1,Groups B, C and D. T1 to T4
Type X and Y Pressurization for Class I, Division 1, Groups B, C and D. T1 to T4
ATEX:  II2G Ex d px IIB+H2 T1...T4 Gb
IECEx:  Ex d px IIB+H2 T1...T4 Gb
TIIS:     Ex pd IIB+H2 T1 to T4

Protection degree of enclosure

NEMA3R, Equivalent to IP54 (dust and water resistant structure)

Operating ambient
conditions

-10 to 50, 95% RH or less (no condensation)

Weight

Wall-mounting version

Self-standing version

Type 1

 approx. 100 kg 

 approx. 140 kg 

Type 2

approx. 155 kg

approx. 190 kg

Type 3

approx. 200 kg

approx. 220 kg

Isothermal Oven

Volume

Large isothermal oven: Approximately 45 L
Standard isothermal oven: Approximately 31 L

Setting temperature
range

55 to 225
(Temperature can be set in one-degree step.)

Temperature control
accuracy

± 0.03

Temperature control

PID control

Input/Output

Analog Input/Output

Maximum of 16/Maximum of 32

Contact Input/Output

Maximum of 32/Maximum of 20

PC communication

Ethernet communication
Protocol: TCP/IP, FTP, Modbus TCP/IP

DCS communication

RS-422
Protocol: MODBUS, Y-Protocol


Utility

Power supply

100/110/115/120/200/220/230/240 V AC ±10%,
50/60 Hz ±5%

Maximum rated power

Type 1: 0.8 to 1.6 kVA
Type 2: 1.4 to 2.9 kVA
Type 3: 2.0 to 4.3 kVA

Instrument air

Pressure: 350 to 900kPa
Flowrate:
   Type 1:                 100 to 140 L/min
      Type1 with FPD: 130 to 200 L/min
   Type 2:                 150 to 210 L/min
      Type2 with FPD: 180 to 270 L/min
   Type 3:                  200 to 280 L/min

Carrier gas

Types:

H2, N2, He, or Ar

Purity:
   Measuring range from 0 to 50 ppm or more:

99.99% minimum (water: 10 ppm or less, organic components: 5 ppm or less)

   Measuring range from 0 to less than 50 ppm:

99.999% minimum (water: 5 ppm or less, organic components: 0.1 ppm or less)

Pressure:

H2: 500 kPa (72.5 psi) (Supplied with extra-regulator for explosion-proof certification)
Other than H2: 400 to 700 kPa

Consumption:

60 to 300 mL/min per isothermal oven

#
copyright 2009-2021  HT INDUSTRIAL AUTOMATION LIMITED