YOKOGAWA OX400 Low Concentration Zirconia Oxygen Analyzer

YOKOGAWA OX400 Low Concentration Zirconia Oxygen Analyzer

This is the latest oxygen analyzer from Yokogawa, and its development was based on the company's long experience and strong track record with this technology.
A proprietary new thin-film deposition technology was used in the zirconia sensor that creates a molecular bond between the zirconia element and the platinum layer. This prevents separation, enables a reduction in sensor size and ensures a high-speed response and long life.
The OX400 can be used to control and monitor various semiconductor applications, and to control environment, air leakage into inert gas, and other processes.

Oxygen concentration control in semiconductor-related diffusion and drying furnaces and in LCD manufacturing processes

Oxygen concentration control in solder pot flow and re-flow ovens, and glove boxes used in electronics manufacturing, and in gas production processes

Oxygen concentration measurements to prevent dust explosions during powder transfer

 Bulletin for OX400

 General Specification for OX400

 User's Manual for OX400

 Selection Guide for Oxygen Analyzer

Features

Long life and high-speed response

Thanks to the use of Yokogawa's proprietary new thin-film deposition technology, the sensor has three times the lifespan of those used in our earlier products

A cylindrical sensor design facilitates the replacement of measurement gases, thereby helping to assure a high-speed response

High performance and high reliability

Superior repeatability and linearity even at low oxygen concentrations

Either pump or aspirator sampling can be selected, depending on the application

Built-in functions and a variety of self-diagnosis functions

Comes with multi selector, auto range, partial range, and pump on/off functions

A variety of self-diagnosis functions are provided that detects malfunctions such as heater temperature error, temperature sensor burnout, and sensor resistance value error.

Superior maintainability

The sensor can be replaced on-site

Compact and lightweight for easy installation

Specifications

Measurement object

Oxygen concentrations in inert gases containing no flammable gas, silica, corrosive gas, or liquid (including water vapor)

Measurement system

Zirconia system

Sampling method

Pump, aspirator, or no suction device

Sample gas conditions

Flow rate: 200 ± 25 ml/min (only applies to sensor)
Temperature: 0-50
(non-condensing)
Humidity: Non-condensing
Pressure: 0-300 PaG
Max. tolerance pressure: 30 kPaG

Measurement range

0-10 ppm O2 to 0-100 vol% O2

Output range

Auto:

0-10 ppm, 0-100 ppm, 0-1000 ppm, 0-1%, 0-10%, 0-100% (default) Other: 0- 0 ppm,0- 00 ppm, 0- 000 ppm, 0- %, 0- 0%, 0-100% can be set, is an integer from 1 to 9

Fixed:

Set to 0-10 ppm, 0-100 ppm, 0-1000 ppm, 0-1%, 0-10%, or 0-100%

Partial:

Lower value or upper value of range can be set

Analog output

2 outputs,
Primary: 4 to 20 mA DC (maximum load resistance: 550Ω )
Secondary: Select from 0-1, 0-5, 0-10 V DC (load resistance: 10 kΩ or greater)

Contact output

3 points :

error O2 , concentration alarm contact, Range marker output contact

Multi selector(optional) :

Contact output for switching of measurement flow path, contact output for measurement flow path data

Contact input

1 point; remote switching for sample gas suction pump ON/OFF

Self-diagnostics

Error (failure):

Sensor unit error, heater temperature error, temperature sensor disconnection, device temperature error, CPU error, fan stop

Alarm (warning):

Heater unstable, sensor defect, electromotive force abnormal, asymmetry voltage error, calibration error, sensor resistance error, O2 concentration upper/lower, over range

Serial communication

RS-232,38400 bps

Calibration methods

(1) 3 point: 10 ppm, 1000 ppm, air, (2) 2 point: zero and span calibration may be set freely, (3) 1 point, (4) Air calibration

Warm-up time

Within 20 min

Power supply

100 - 120 V AC/200 - 240 V AC, 50/60 Hz

Power consumption

100 to 120 V AC, 200 VA max., 200 to 240 V AC, 400 VA max

Dimension

213 (W) x 132 (H) x approx. 375 (D) mm

Weight: Approx.

5 kg

Line connection

gas inlet: Rc1/4 or 1/4NPT female
gas outlet: Rc1/4 or 1/4NPT female

Environment and operational conditions

Installation conditions:

Indoors, panel or wall mounting, non explosion area

Ambient temperature:

0 to 40, non-condensing

Ambient humidity:

5 to 85% RH

Storage temperature:

-5 to 50

Conformance to safety and EMC standards

Safety:

61010-1:2001

EMC:

EN 61326-1 Class A, Table 2 (For use in industrial locations) EN 61326-2-3, EN 61000-3-2, EN 61000-3-3

Characteristics

Repeatability

±1% FS (Hereafter, either 10, 100, 1000 ppm, 1%, 10%, or 100% O2 is FS)

Linearity

±2% FS
±3% FS (0-100 ppm or less)

Response time

90% response
within 10 sec (0-1% or more)
within 30 sec (less than 0-1%)

Drift

±2% FS / week

#