YOKOGAWA ZR202S |
YOKOGAWA ZR202S
Measurement object |
Oxygen concentration in combustion exhaust gas and mixed gas. |
Measurement system |
Zirconia system |
Display |
6-digit LCD |
Switch |
Three optical switches |
Explosionproof Approval |
ZR202S-A (ATEX); EExd II B + H2, Group II, Category 2GD, T2, T300℃ ZR202S-B (FM); Class I, Division1, Groups B, C and D, Class II/III, Division1, Groups E, F and G, T2 ZR202S-C (CSA); Class I, Division1, Groups B, C and D, Class II/III, Division1, Groups E, F and G, T2 ZR202S-D (IECEx); Exd II B + H2, T2, ExtD A21 IP66 T300℃ |
Oxygen concentration |
0.01 to 100 vol% O2 |
Output signal |
4 to 20 mA DC, one point (maximum load resistance 550Ω) |
Measurement range |
Any setting in the range of 0 - 5 vol%O2 to 0 - 100 vol%O2 (in 1 vol% O2), or partial range. |
Digital communication (HART) |
250 to 550Ω, depending on number of field devices connected to the loop (multi-drop mode) |
Sample gas pressure |
- 5 to + 5 kPa No pressure fluctuation in the furnace should be allowed. |
Sample gas temperature |
0 to 700℃ |
Probe length |
0.4, 0.7, 1.0, 1.5, 2.0 m |
Contact output |
Two points (one is fail-safe, normally open) |
Contact input |
Two points |
Calibration |
Method; zero/span calibration Calibration mode; automatic, semi-automatic and manual (with optical switch) |
Construction |
Equivalent to NEMA4X/IP66 |
Ambient temperature |
-20 to +55℃ (- 5 to +70℃ on the case surface) |
Power Supply |
85 to 264 V AC, 45 to 66 Hz |
Performance
Repeatability |
±0.5% Maximum value of set range |
Linearity |
±1% Maximum value of set range |
Drift |
±2% Maximum value of set range/month |
Response time |
90% response within 5 sec | |
|