HT INDUSTRIAL AUTOMATION LIMITED

 
YOKOGAWA FC500G Free Available Chlorine Analyzer

YOKOGAWA FC500G Free Available Chlorine Analyzer

FC500G non-reagent free available chlorine analyzer adopts the polarographic method, using a rotating electrode to continuously measure online the concentration of free available chlorine.

The FC500G is intended for measurements in the water distribution network, at pumping stations, water delivery points, schools, and public facilities.

 General Specification for FC500G

 User's Manual for FC500G

Features

*

Measuring range can be selected from 0 to 1/0 to 2/0 to 3mg/l and 0 to 5/0 to 10 mg/l.

*

Rotating electrode/ceramic bead cleaning has been improved to making cleaning automatic and continuous.

*

Easy-to-maintain (-clean) cell construction.

Specifications

Measured Object

Free chlorine in tap water

Measuring System

Polarographic method using rotating electrode

Measuring Range

Any of 0 to 1/0 to 2/0 to 3 mg/l and 0 to 5/0 to 10 mg/l certainly, use to air purge

Output Signal

4 to 20mA DC (load resistance: Maximum 550 Ω)

Sample Solution Condition

Temperature: 0 to 40
pH: 6.5 to 7.5
Flow Rate: 0.1 to 2.5 l/min (at cell inlet)
Pressure: 1 to 30 kPa (without VALVE)
              1 to 150 kPa (with VALVE)
Conductivity: 100 to 300 μS/cm
SS (Suspended Solid): 10 mg/l or less

Electrode

Indicating electrode: Rotating gold electrode
Reference electrode: Silver electrode (with Pt1000 RTD installed)

Electrode Cleaning

Ceramic bead cleaning

Ambient Temperature

-10 to 50

Installation

Indoor

Power Supply

100 V AC/110 V AC / 200 V AC/220 V AC ±10 %, 50/60 Hz

Weight

Approx. 6.5 kg

Performance

Linearity

±5 % of full scale

Repeatability

2 % of full scale

Stability

Zero drift; ±1 % of full scale/month or less
Span drift; -10 % of full scale/month or less

Response time

Approx. 2 min (90 % response time)

#
copyright 2009-2021  HT INDUSTRIAL AUTOMATION LIMITED