HT INDUSTRIAL AUTOMATION LIMITED

 
YOKOGAWA FC400G Free Available Chlorine Analyzer

YOKOGAWA FC400G Free Available Chlorine Analyzer

The FC400G non-reagent free available chlorine analyzer adopts the polarographic method, using a rotating electrode to continuously measure online the concentration of free available chlorine.

The FC400G is a microprocessor-based, intelligent free chlorine analyzer whose development was based on Yokogawa's long experience with residual chlorine analysis.

The FC400G is intended for measurements in the water distribution network, at pumping stations, water delivery points, schools, and public facilities.

 Bulletin for FC400G

 General Specification for FC400G

 User's Manual for FC400G

 Click here to request a technical document for Chlorine Analyzer

Features

*

One-touch simple calibration.

*

Output range can be arbitrarily set in the range of from 0-1 mg/l to 0-3 mg/l in the field.

*

Line segment output is available for output signals.

*

Easy confirmation of inter-electrode applied voltage/current characteristic (plateau characteristic). This allows understanding of the detailed sensor characteristics in the application.

*

Rotating electrode/ceramic bead cleaning has been improved to realize further effective and continuous automatic cleaning.

*

Easy-to-maintain (-clean) cell construction.

Specifications

Measured Object

Free chlorine in tap water

Measuring System

Polarographic method using rotating electrode

Measuring Range

0 to 3 mg/l

Output Range

Can switch between arbitrary ranges (each between 0 to 1 mg/l and 0 to 3 mg/l)

Output Signal

4 to 20mA DC (load resistance: 550 Ω or less) or 1 to 5V DC (output resistance : 300 Ω or less)

Sample Conditions

Temperature: 0 to 50
pH: 6.5 to 7.5
Flow Rate: 0.1 to 2.5 l/min
Pressure: 1 to 150 kPa
          100 to 740 kPa (when a dedicated sampling unit is provided)
Conductivity: 100 to 300 μS/cm
SS (Suspended Solid): 10 mg/l or less

Electrode

Indicating electrode: Rotating gold electrode
Reference electrode: Silver electrode (with Pt1000 RTD installed)

Electrode Cleaning

Ceramic bead cleaning

Ambient Temperature

-30 to 70

Installation

Indoor (Outdoor use requires protection against rain.) (Direct sunlight must be avoided.)

Power Supply

100/110/200/220 V AC ±10 %, 50/60 Hz

Weight

Approx. 6 kg

Performance

% display shows value relative to the upper limit of a range.

Linearity

±5 % of full scale

Repeatability

2 % of full scale

Stability

Zero drift; ±1 % of full scale/month or less
Span drift; -10 % of full scale/month or less

Response time

Approx. 2 min (90 % response time)

#
copyright 2009-2021  HT INDUSTRIAL AUTOMATION LIMITED