HT INDUSTRIAL AUTOMATION LIMITED

 
YOKOGAWA NR801EL Near Infrared Analyzer

YOKOGAWA NR801EL Near Infrared Analyzer

The InfraSpec NR801 EL is the at-line model of NR800 series FT-NIR analyzer. It employs exactly the same hardware in the heart of the system (interferometer and detector) as NR800 process model and thus realizing high S/N (signal-to-noise) ratio, high wave number resolution and wide wavenumber scanning range those done in the NR800 process model. Direct model transfer to/from the NR800 series including process model is another key feature. This model transfer capability will drastically cut down model implementation time and cost at project stage while it provides more flexible and efficient model upgrade during routine operation.

Software specially developed for at-line model, SpectLand 2 at-line, will make operation simple and user-friendly and will improve the work efficiency for at-line and Lab. measurement application.

 Bulletin for NR801EL

 General Specification  for NR801EL

 Click here to request a technical document for Near Infrared Analyzers

 

Features

*

Newly developed high-performance interferometer and detector.

*

Calibration model transfer and share between NR800 series.

*

Various types of measurement cells (cuvette, vial, probe, and flow-through, etc.,).

*

High-speed Ethernet communication between PC and the analyzer.

*

At-line analysis software (SpectLand 2 at-line) for simple operation.

Specifications

Location

Non-hazardous area, indoor.
Avoid exposure to wind and rain, sunlight or radiation heat.

Principle

Fourier-transform Near Infrared analyzer via fiber-optic cable, Double beam

Measurement method

Optical transmission absorption

Detector

InGaAs (Indium Gallium Arsenide)

Light source

Halogen lamp

Wavelength range

900 to 2500 nm

Wavelength accuracy

0.04 cm-1

Wavelength repeatability

0.007 cm-1

Wavelength resolution

4, 8, 16, 32, 64 cm-1 selectable

Number of channel

1 channel

Number of measurement

Max. 12 per analyzer

Measurement time

4 seconds as minimum(depending on the number of averaging and measuring channels, as well as measurement items)

Communication interface

Engineering PC: 1 channel (Ethernet)

Contact output

Two points; Purge failure, analyzer failure

Enclosure

General purpose type:
  Model NR801EL; with CE mark
  Model NR801JL; Japanese version

Power supply

100, 115, 200 or 230 V AC ±10 %, 50/60 Hz; consumption:approx. 200 VA

Environment temp., RH

0 to 35, 0 to 85 % RH

Weight

Approx. 35 kg

Performance

Wavelength reproducibility

0.007 cm-1

Wavelength accuracy

0.04 cm-1

S/N ratio

2250:1 (RSM, 4 cm-1 resolution, 4100 to 4200 cm-1, 1 sec.)

#