YOKOGAWA ISC450G 4-Wire Inductive Conductivity Analyzer

YOKOGAWA ISC450G 4-Wire Inductive Conductivity Analyzer

The EXAxt 450 will help you and guide you as the unique Human Machine Interface (HMI) will not only present you reliable process data, but diagnostics, trends, logbooks and step by step calibration routines to become your friend in analytical measurement and help you gaining that higher level of quality. The high resolution graphical display and the touchscreen operation make all information visible to the operator.

The ISC450G is designed to measure specific conductivity by using the inductive conductivity method. This method is especially suitable for high conductivity measurements. The ISC450G features a unique accuracy and rangeability, which is only possible through superior sensor design and electronic data processing using a self tuning preamplifier with more than 10 sub ranges with automatic zero adjustment at every range change.

Accurate concentration measurement requires accurate temperature compensation and the ISC450G offers three compensation methods from default NaCl compensation to user defined MATRIX compensation for electrolytes where the coefficient varies with temperature and with concentration.

 Bulletin for ISC450G 

 General Specification for ISC450G 

 General Specification for Conductivity Sensors 

 User's Manual for ISC450G 

 Selection Guide for Conductivity Analyzer 

Features

*

Unique intuitive HMI menu structrure in 8 languages.

*

Process data trending up to 2 weeks.

*

Predefined OIML standard solutions.

*

Predefined matrices for precise temperature compensation.

*

Two mA-outputs and four SPDT relay contacts with display indicators.

*

% weight on display.

*

HART® Communications.

*

FM Class I, Div.2 Group A, B, C and D. T6 for Ta -20 to 55.

Specifications

Input specifications

Compatible with ISC40 inductive conductivity sensor with integrated temperature sensor: NTC30k or Pt1000.

Input range

Conductivity: 0 to 1999 mS/cm at 25 (77)
  Minimum: 1 μS/cm (at process temperature)
  Maximum: 2 S/cm (at process temperature)
Temperature: -20 to 140
(0 to 280).

Cable length

Max. 60 meters (200 feet). 10 meters (35 feet) fixed sensor cable + 50 meters (165 feet) WF10 extension cable.

Transmission signal

Two isolated outputs of 4 to 20 mA DC with common negative. Maximum load 600 Ω. Bi-directional HART® digital communication, superimposed on mA1 (4 to 20 mA) signal. Burn up (21.0 mA) or burn down (3.6 mA) to signal failure accorded to NAMUR NE43.

Contact output

Four SPDT relay contacts with display indicators.

Contact input

Remote range switching to 10 times the programmed range.

Temperature compensation

Automatic or manual, for temperature ranges.
 Reference temp. programmable from 0 to 100
or 30 to 210 (default 25).

Calibration

Semi-automatic calibration using pre-configured OIML (KCl) stanard tables, with automatic stability check. Manual adjustment to grab sample.

Display

Graphical Quarter VGA (320 x 240 pixels) LCD with LED backlight and touch screen.

Housing

Cast Aluminim housing with chemically resistant coating; Polycarbonate cover with Polycarbonate flexible window
Protection IP66/NEMA 4X/CSA Type 3S

Environment

Ambient Temperature: -20 to 55
Humidity: 10 to 90 % RH at 40
(non-condensing)

Power supply

ISC450G-A:
  Ratings; 100 to 240 V AC
  Acceptable range; 90 to 264 V AC
  Ratings; 50/60 Hz
  Acceptable range; 50 Hz ±5%, 60 Hz ±5%
  Power Consumption; 15 VA
ISC450G-D:
  Ratings; 12 to 24 V DC
  Acceptable range; 10.8 to 26.4 V DC
  Power Consumption; 10 W

Performance

Conductivity

≤0.5 % of reading ± 1.0 μS/cm

Temperature

≤0.3 (0.6)

Step response

≤4 seconds for 90 % (for a 2 decade step)

#