YOKOGAWA HANDY MULTIFUNCTION CALIBRATOR CA71
Handy Multifunction Calibrator Signal source: DCV, DCA, Resistance, RTD, TC (10 kinds), Frequency, Pulse Measurement: DCV, DCA, ACV, Resistance, RTD, TC (10 kinds), Frequency, Pulse Features: Simultaneous signal source and measurement capacity, a wide array of useful functions, Easy operation, Online communication function Options: AC adpeter (94012: For 100V AC, 94013: For 120V AC, B9108WB: For 220-240V AC), B9108WA RJ sensor, B9108XA Accessory carrying case, 91017 Communications cable
Specification - Model CA71
|
Auto-Step Output Function |
n value sent automatically when n/m function selection is selected (two options: approximately 2.5 seconds/step or approximately 5 seconds/step) |
|
|
Sweep Functions |
Sweep time (two options: approximately 16 seconds or approximately 32 seconds) |
|
|
Memory Function |
50 value sets (generated and measured values are stored as value sets with the same address (up to 50 value sets can be stored)) |
|
|
Measuring Unit Maximum Input |
Voltage terminal: 300 V AC Current terminal: 120 mA DC |
|
|
Current Terminal Input Protection |
Fuses: 125 mA/250 V |
|
|
Measuring Unit Ground Voltage |
Maximum 300 V AC |
|
|
Measurement Display Updating Rate |
Approximately once per second |
|
|
Serial Interface |
Enabled when communication cable (RS232) is connected; sold separately as optional accessory |
|
|
Display |
Segmented LCD (Approx. 76 mm × 48 mm) |
|
|
Backlight |
LED backlight; auto-off after one minute (from when LIGHT key is turned on) |
|
|
Power Supply |
Four AA alkaline batteries, or special AC adapter (sold separately) |
|
|
Battery Life |
Measurement off, output 5 V DC/10 kO or greater: Approximately 40 hours Simultaneous signal generation/measurement, output 5 V DC/10 kO or greater: Approximately 20 hoursSimultaneous signal generation/measurement, output 20 mA/5 V: Approximately 12 hours (using alkaline batteries, with backlight off) |
|
|
Auto-Power-Off Function |
Approx. 10 minutes (auto-power-off can be disabled through a DIP switch setting) |
|
|
Applicable Standard |
EN55011: 1998, Class B, Group 1 EN61326-1: 1997 + A1: 1998 IEC61010-1, IEC61010-2-31 |
|
|
Insulation Resistance |
Across input terminal and output terminal, 500 V DC, 50 MO or greater |
|
|
Withstand Voltage |
Across input terminal and output terminal, 3.7 kVAC, for one minute |
|
|
Operating Temperature And Humidity Ranges |
0 - 50°C, 20 - 80% RH (no condensation) |
|
|
Storage Temperature And Humidity Ranges |
-20 - 50°C, 90% RH or less (no condensation) |
|
|
External Dimensions |
Approx. 190 × 120 × 55 mm |
|
|
Weight |
Approx. 730 g (including batteries) |
Product Overview
|
Easy Operation |
The Handy Calibrator incorporates rotary switches for simple handling. Just open the carrying case cover and connect the cables, and you're ready to take measurements. |
|
|
A Wide Array of Useful Functions |
- Source
Values set in steps of 4-20 mA 24V DC Power Supply to Transmitter
- Divided output (n/m) function
Output settings are divided, eliminating the need for bothersome calculations for percentage output.
- Autostep function
Changes the output value in step form based on the setting from the divided output (n/m) function. Changes can be sourced automatically every 10% or 25%.
- Online communication (CA71 only) RS-232C-compliant optically isolated interface
- Sweep function
Linearly increases or decrease the output. The increasing/decreasing time can be set to either 16 or 32 seconds.
- Memory function
Source values and measurements forming individual value sets can be saved to or read from the Handy Calibrator's internal memory (maximum 50 value sets).
- Temperature monitor function
|
Parts & Acessories
|
Supplied Accessories |
CA71/SP1 includes Communications cable, data logging softare, 120V power supply, magnetic backing instrument case, source/measure cables, NIST traceable certification. |
|
|
Standard Accessories |
Carrying case: 93016 Four AA alkaline batteries: A1070EB × 4 Fuse: A1501EF (for current terminal input protection) Measurement lead cables (one red, one black): RD031 Signal generating lead cables (one red, two black): 98020 Terminal adapter: 99021 User’s manual: IM CA71-E |
|
|
Optional Accessories (Sold Separately) |
AC adapter: 94012 (100 V AC power supply) AC adapter: 94013 (120 V AC power supply) AC adapter: B9108WB (220–240 V AC power supply) Accessory carrying case: B9108XA Communication cable: 91017 RJ sensor: B9108WA (For reference junction compensation) |
|
|
Spare Parts |
Carrying case: 93016 Fuse: A1501EF (for current terminal input protection) Measurement lead cables (one red, one black): RD031 Signal generating lead cables (one red, two black): 98020 Terminal adapter: 99021 | |