YOKOGAWA HANDY CALIBRATOR CA11E
Voltage Calibrator - Current Calibrator
Signal source: DCV (Max. 30V), DCA (Max. 24 mA)
Measurement: DCV (Max. 30V), DCA (Max. 100 mA)
Features: Auto step (4 to 20mA), 20mA SINK, Sweep function
Options: CA11E-2 Includes rubber boot, strap and accessory case
Specifications - Model CA11E-1
|
Features |
Source and Measuring of Voltage and Current Generates and measures voltages up to 30 V DC and currents up to 24 mA DC.
Improved display resolution for current The generation/measurement resolution has improved to 0.001 mA.
24 V (20 mA)/Loop check function Enables measurement of generated current signals while supplying loop power to two-wire transmitter. Load current: Max. DC22 A
20 mA SINK Function Absorbs the voltage supplied from an external power supply to its H terminal and simulates a two-wire transmitter making it ideal for loop checks.
4-20 mA and 1-5 V DC Step-up/down Function The output level can be changed between 4-8-12-16-20 mA signals by one touch for the 4-20 mA DC output, and between 1-2-3-4-5V signals for the 1-5 V DC output, for efficient calibration work.
Sweep Function Increases and decreases the output level to the preset level at a constant rate for the selected sweep time (16 or 32 seconds). The sweep function and sweep time are set by the internal dip switches. |
|
|
Standard Accessories |
Carrying Case |
|
|
Power Supply |
Four 1.5-V alkaline batteries (ANSI AA-size) or dedicated AC adapter (sold separately) |
|
|
Battery Life |
Approx. 25 hours for 20 mA DC output (with a load of 5 V) When generating a signal continuously on alkaline batteries Approx. 50 hours for 5 V DC output (with a load of 10k? or greater) When generating a signal continuously on alkaline batteries |
|
|
Automatic Power-Off |
Approx. 10 minutes (Can be canceled by DIP switch setting) |
|
|
Generation Signal Level Setting |
By four-digit up/down keys |
|
|
Response of Generator |
Approx. 1 second (from when the output begins to change until when the output level falls within the specified accuracy) |
|
|
Maximum Allowable |
30 VDC or less between each terminal and ground |
|
|
Measurement Display |
Approx. 1 update/sec. |
|
|
Update interval Display |
7 segments LCD |
|
|
Operating Temperature/Humidity Range |
0 to 50°C and 20 to 85%RH (no condensation) |
|
|
Storage Temperature/Humidity Range |
-20 to 50°C and 90%RH or less (no condensation) |
|
|
Dimensions |
Approx. 192 (W) x 92 (H) x 42 x (D) mm (Excluding protrusions) |
|
|
Width |
192 mm |
|
|
Height |
92 mm |
|
|
Thickness |
42 mm |
|
|
Weight |
Approx. 440 g (including batteries) |
|
|
Additional Features |
Complete Protection (CA11E) Dip Switches (inside battery compartment) Longer Lead Cables Rubber Boot Runs on 1.5 V AA-size Batteries or AC Adapter Simple and Easy Calibration |
Additional Features
Choose either "rubber boot" or "carrying case"
Rubber boot, strap and accessory case can be chosen as standard accessories to improve impact resistance and make it easier to carry. (The calibrator cannot be stored in the carrying case when the rubber boot is attached.)
The roomy case easily holds the calibrator without having to disconnect the lead cables.
Automatic Power-off
Power is turned off automatically if the calibrator is not touched for 10 minutes, prolonging battery life.
Runs on 1.5 V AA-size Batteries or AC Adapter
The handy calibrators can run on the built-in 1.5-V ANSI batteries or an AC power supply using the AC adapter (sold separately).
Longer Lead Cables
The slightly longer than usual lead cables of 1.7 m (approximately 0.1 Ω for both cables) allow easy cable connection even if the handy calibrator is put on the floor.
Compact and Lightweight
Almost the same size and weight as a hand-held digital multimeter, this calibrator is designed for use in the field.
Simple and Easy Calibration
For the CA11E and CA12E, There is performing calibrations during periodic maintenance as they can be simply calibrated by the up/down keys.
Complete Protection (CA11E)
The complete protector protects the circuit against short-circuiting of the voltage output terminals and application of a voltage (of up to 30 V) to the output terminals, etc. due to misconnection.
Choose either "rubber boot" or "carrying case"
Rubber boot, strap and accessory case can be chosen as standard accessories to improve impact resistance and make it easier to carry. (The calibrator cannot be stored in the carrying case when the rubber boot is attached.)
The roomy case easily holds the calibrator without having to disconnect the lead cables.
Automatic Power-off
Power is turned off automatically if the calibrator is not touched for 10 minutes, prolonging battery life.
Runs on 1.5 V AA-size Batteries or AC Adapter
The handy calibrators can run on the built-in 1.5-V ANSI batteries or an AC power supply using the AC adapter (sold separately).
Longer Lead Cables
The slightly longer than usual lead cables of 1.7 m (approximately 0.1 Ω for both cables) allow easy cable connection even if the handy calibrator is put on the floor.
Compact and Lightweight
Almost the same size and weight as a hand-held digital multimeter, this calibrator is designed for use in the field.
Simple and Easy Calibration
For the CA11E and CA12E, There is performing calibrations during periodic maintenance as they can be simply calibrated by the up/down keys.
Complete Protection (CA11E)
The complete protector protects the circuit against short-circuiting of the voltage output terminals and application of a voltage (of up to 30 V) to the output terminals, etc. due to misconnection.
Dip Switches (inside battery compartment)
|
CA11E |
CA12E |
Switch Number |
ON |
OFF |
ON |
OFF |
1 |
Sweep function |
Sweep function -off |
The built-in RJC is on. |
External RJC is on |
2 |
32 s |
16 s |
°F |
°C |
3 |
24 |
24 |
PT100-IPTS68 |
PT100-ITS90 |
4 |
Automatic power-off is disabled. |
Automatic power-off |
Automatic power-off is disabled. |
Automatic power-off |
Specifications - Model CA11E-2
|
Features |
Source and Measuring of Voltage and Current Generates and measures voltages up to 30 V DC and currents up to 24 mA DC.
Improved display resolution for current The generation/measurement resolution has improved to 0.001 mA.
24 V (20 mA)/Loop check function Enables measurement of generated current signals while supplying loop power to two-wire transmitter. Load current: Max. DC22 A
20 mA SINK Function Absorbs the voltage supplied from an external power supply to its H terminal and simulates a two-wire transmitter making it ideal for loop hecks.
4-20 mA and 1-5 V DC Step-up/down Function The output level can be changed between 4-8-12-16-20 mA signals by one touch for the 4-20 mA DC output, and between 1-2-3-4-5V signals for the 1-5 V DC output, for efficient calibration work.
Sweep Function Increases and decreases the output level to the preset level at a constant rate for the selected sweep time (16 or 32 seconds). The sweep function and sweep time are set by the internal dip switches. |
|
|
Standard Accessories |
Accessory case Rubber Boot Strap |
|
|
Power Supply |
Four 1.5-V alkaline batteries (ANSI AA-size) or dedicated AC adapter (sold separately) |
|
|
Battery Life |
Approx. 25 hours for 20 mA DC output (with a load of 5 V) When generating a signal continuously on alkaline batteries Approx. 50 hours for 5 V DC output (with a load of 10k? or greater) When generating a signal continuously on alkaline batteries |
|
|
Automatic Power-Off |
Approx. 10 minutes (Can be canceled by DIP switch setting) |
|
|
Generation Signal Level Setting |
By four-digit up/down keys |
|
|
Response of Generator |
Approx. 1 second (from when the output begins to change until when the output level falls within the specified accuracy) |
|
|
Maximum Allowable |
30 VDC or less between each terminal and ground |
|
|
Measurement Display |
Approx. 1 update/sec. |
|
|
Update interval Display |
7 segments LCD |
|
|
Operating Temperature/Humidity Range |
0 to 50°C and 20 to 85%RH (no condensation) |
|
|
Storage Temperature/Humidity Range |
-20 to 50°C and 90%RH or less (no condensation) |
|
|
Dimensions |
Approx. 192 (W) x 92 (H) x 42 x (D) mm (Excluding protrusions) |
|
|
Width |
192 mm |
|
|
Height |
92 mm |
|
|
Thickness |
42 mm |
|
|
Weight |
Approx. 440 g (including batteries) |
|
|
Additional Features |
Carrying Case Complete Protection (CA11E) Dip Switches (inside battery compartment) Longer Lead Cables Runs on 1.5 V AA-size Batteries or AC Adapter Simple and Easy Calibration |
Additional Features
Choose either "rubber boot" or "carrying case"
Rubber boot, strap and accessory case can be chosen as standard accessories to improve impact resistance and make it easier to carry. (The calibrator cannot be stored in the carrying case when the rubber boot is attached.)
The roomy case easily holds the calibrator without having to disconnect the lead cables.
Automatic Power-off
Power is turned off automatically if the calibrator is not touched for 10 minutes, prolonging battery life.
Runs on 1.5 V AA-size Batteries or AC Adapter
The handy calibrators can run on the built-in 1.5-V ANSI batteries or an AC power supply using the AC adapter (sold separately).
Longer Lead Cables
The slightly longer than usual lead cables of 1.7 m (approximately 0.1 Ω for both cables) allow easy cable connection even if the handy calibrator is put on the floor.
Compact and Lightweight
Almost the same size and weight as a hand-held digital multimeter, this calibrator is designed for use in the field.
Simple and Easy Calibration
For the CA11E and CA12E, There is performing calibrations during periodic maintenance as they can be simply calibrated by the up/down keys.
Complete Protection (CA11E)
The complete protector protects the circuit against short-circuiting of the voltage output terminals and application of a voltage (of up to 30 V) to the output terminals, etc. due to misconnection.
Panel Designs Common to Both Models
Power Switch
Up/Down Keys
Used to set the output signal level. A pair of up/down keys is conveniently located immediately below each digit in the LCD panel.
Output On/Off Switch
(Output signal can set ON/OFF)
Range Selection Rotary Switch
The rotary switch simplifies range selection: just leave the switch set to the most frequently used range.
Dip Switches (inside battery compartment)
|
CA11E |
CA12E |
Switch Number |
ON |
OFF |
ON |
OFF |
1 |
Sweep function |
Sweep function -off |
The built-in RJC is on. |
External RJC is on |
2 |
32 s |
16 s |
°F |
°C |
3 |
24 |
24 |
PT100-IPTS68 |
PT100-ITS90 |
|
|
|
|
|
4 |
Automatic power-off is disabled. |
Automatic power-off |
Automatic power-off is disabled. |
Automatic power |